Heart Stream by Agilent AED Trainer

 

Buy It 

 

AED MODELS

 

WITH ECG        M3860A

NO ECG                       M3861A

 

DEFIBRILLATOR

 

HOW SUPPLIED                                   Defibrillator, User’s Guide, Battery Pack,  Electrodes (2 pair), Quick Reference Card

OPERATING MODE                               Semi-automatic

 

WAVEFORM                                         Truncated Exponential Biphasic.  Waveform Parameters adjusted as a function of patient impedance.

ENERGY                                              Single energy output.

                                                            Nominal:  150 Joules into a 50ohm. Load

CHARGE TIME FROM SHOCK               Typically less than 10 seconds

ADVISED

 

SHOCK-TO-SHOCK CYCLE                   Typically less than 20 seconds (including analysis

TIME                                                     time) in semi-automatic mode.

                                                           

VOICE AND TEXT PROMPTS                 Extensive text prompts and audible messages guide user through protocol.

PROTOCOL                                          Follows pre-configured settings.  Can be modified with the M3864A Training & Administration Pack.

SHOCK DELIVERY                                Via defibrillation pads placed in anterior-anterior (lead II) position

CONTROLS                                           On/off, Shock, screen contrast/option buttons

    (Standard on both models)

INDICATORS                                         LCD screen, beeper, audio speaker, status indicator, shock button, connector socket LED

ADVANCED MODES                             Configurable

 

PHYSICAL

 

SIZE:          HEIGHT                              6.6 cm. (2.6 inches)

                    WIDTH                              21.8 cm. (8.6 inches)

                    DEPTH                             21.8 cm. (8.6 inches)

WEIGHT   (With battery)                         2.1 kg. (4.7 pounds)

                 (Without battery)                   1.8 kg. (3.9 pounds)

 

PATIENT ANALYSIS SYSTEM

 

PATIENT ANALYSIS                              Per Protocol, evaluates patient ECG and signal quality to determine if a shock is appropriate, and evaluates connection impedance for proper defibrillation pad contact.

SENSITIVITY / SPECIFICITY                  Meets AAMI guidelines.

 

BATTERY PACK (M3863A)

 

TYPE                                                    12 VDC 4.2 Ah lithium manganese.  Disposable, recyclable, long-life, primary cells.

CAPACITY                                            Minimum 300 shocks or 12 hours operating time.

INSTALL-BY-DATE                                Battery is labeled with an install by date at least 5 years from date of manufacture.

STANDBY LIFE                                     Defines how long the battery will power the AED I standby operation within the standby temperature range (one battery insert test and no uses).  4 years minimum when battery is installed by the Install-By-Date.  5 years typical.

 

ECG DISPLAY (M3860A)

 

SCREEN                                               High Resolution LCD with bright backlight.

SCREEN DIMENSIONS                         2.8 inches wide x 2.3 inches high (7.0 cm x 5.8 cm)

DISPLAY RANGE                                  Differential:+/- 2 mV full scale (nominal)

SWEEP SPEED                                    23 mm/second (nominal)

FREQUENCY RESPONSE                     1 Hz to 20 Hz (-3dB) (nominal)

SENSITIVITY                                         1.16 cm/mV 9nominal)

HEART RATE                                        30 to 300 beats per minute updated each analysis period during monitoring.

MONITORED LEAD                               Anterior-anterior (lead II) placement with defibrillation pads (M3860A only)

 

DEFIBRILLATION PADS, CABLE, CONNECTOR

 

CONFIGURATION                                  DP2:  Two pair of DP6:  Six pair

HOW SUPPLIED                                   Disposable, self-adhesive defibrillator pads with integrated cable and connector.

SURFACE AREA                                   100 cm2 each (nominal active surface area)

CABLE LENGTH                                    Approximately 122 cm (48 inches)

 

MEDICAL CONTROL/RECORDING FEATURES

 

STANDARD EVENT REVIEW                 Elapsed time and the number of shocks delivered are displayed on screen.

ENHANCED EVENT REVIEW                Optional Data Card expands the above on-screen event review capabilities:  Review chronological events in detail including ECG

 

OPTIONAL PC DATA CARD (M3854A)

                                                            Review the patient's initial ECG on the screen

                                                            4 Hours of event & ECG recording or 30 minutes if Voice Recording is enabled.

 

ENVIRONMENTAL/PHYSICAL REQUIREMENTS

 

SEALING                                              Meets IEC529 class IP54; with battery and data card tray installed

TEMPERATURE                                    Operating:  0-50 Celsius (32-122 F)

HUMIDITY                                             0% to 95% relative humidity (non-condensing)

STANDBY TEMPERATURE                    0-43 Celsius (32-109 F)

STANDBY HUMIDITY                             0% to 75% relative humidity (non-condensing)

                                                            Standby applies to AED with battery installed and stored with defibrillation pads.

ALTITUDE                                             -500 to 15,000 feet per MIL-STD-810E 500.3

                                                            Procedure II

SHOCK/DROP ABUSE                          MIL-STD-810E 516.4 Procedure IV

   TOLERANCE                                      (1 meter, any edge, corner or surface)

VIBRATION                                           MIL-STD-810E 514.4-17

EMI Requirements                                 Tested to EN60601-1-2 Limits.

EMI (Radiated)                                       Method EN55011:  1998 Group 1 Level B.

EMI (Immunity)                                      Method EN61000-4-3:  1998 Level 2.

Aircraft:  Method                                    RTCA/DO-160D:  1997 Section 21 (Category M-Charging).

 

AUTOMATED AND USER ACTIVATED SELF TESTS

 

AUTOMATIC SELF TESTS                     Tests internal circuitry, waveform delivery system, battery capacity. Verifies calibration of key circuits monthly.

AUTOMATIC SELF TEST                       Daily when stored within operating environmental

   FREQUENCY                                     conditions.

STATUS INDICATION                             Dynamic visual and audible indication of self test results.  Indicates device readiness.

BATTERY INSERTION TEST                  Upon battery insertion, extensive automatic self tests and user interactive tests check device readiness.  Verifies calibration of key circuits.

AUTOMATIC STANDBY                         Instrument automatically monitors temperature and

TEMPERATURE MONITORING    warns user if device is stored outside of standby temperature range.

 

TRAINING AND ADMINISTRATION PACK (M3864A)

 

FUNCTION                                            Places FR2 in scenario-based training mode.  10 real world scripts provided.  Disables energy delivery.

                                                            Permits modification of preprogrammed FR2 protocol.

TYPE                                                    12 VDC 1.1 Ah Rechargeable Nickel Metal Hydride

CAPACITY                                            Min 4 hours training time

RECHARGE TIME                                 90 minutes to full capacity using Charger, M3855A (sold separately)

 

CHARGER (M3855A)

 

                                                            For use with the Training and Administration Pack, M3864A.

POWER INPUT/MAINS                          100-240 VAC 47-63 HZ

POWER                                                30 Watts

 

CODERUNNER WEB SOFTWARE

 

OPTIONAL                                            Software for review of data from optional PC Data Cards.

 

Refer to FR2 User's Guide for detailed product use instructions.

All specifications are based on 25 Celsius unless otherwise noted.  Specifications

are subject to change.

The FR2 and accessories are made of latex-free materials.

 

 