Heart Stream by Agilent AED Trainer
AED MODELS
WITH ECG M3860A
NO ECG M3861A
DEFIBRILLATOR
HOW
SUPPLIED Defibrillator,
User’s Guide, Battery Pack, Electrodes
(2 pair), Quick Reference Card
OPERATING
MODE Semi-automatic
WAVEFORM Truncated
Exponential Biphasic. Waveform
Parameters adjusted as a function of patient impedance.
ENERGY Single
energy output.
Nominal: 150 Joules into a 50ohm. Load
CHARGE
TIME FROM SHOCK Typically
less than 10 seconds
ADVISED
SHOCK-TO-SHOCK
CYCLE Typically less
than 20 seconds (including analysis
TIME
time) in semi-automatic mode.
VOICE AND
TEXT PROMPTS Extensive
text prompts and audible messages guide user through protocol.
PROTOCOL Follows
pre-configured settings. Can be
modified with the M3864A Training & Administration Pack.
SHOCK
DELIVERY Via
defibrillation pads placed in anterior-anterior (lead II) position
CONTROLS On/off,
Shock, screen contrast/option buttons
(Standard on both models)
INDICATORS LCD
screen, beeper, audio speaker, status indicator, shock button, connector socket
LED
ADVANCED
MODES Configurable
PHYSICAL
SIZE: HEIGHT 6.6
cm. (2.6 inches)
WIDTH 21.8 cm. (8.6
inches)
DEPTH 21.8 cm. (8.6
inches)
WEIGHT (With battery) 2.1 kg. (4.7 pounds)
(Without battery) 1.8 kg. (3.9 pounds)
PATIENT
ANALYSIS SYSTEM
PATIENT
ANALYSIS Per
Protocol, evaluates patient ECG and signal quality to determine if a shock is
appropriate, and evaluates connection impedance for proper defibrillation pad
contact.
SENSITIVITY
/ SPECIFICITY Meets AAMI guidelines.
BATTERY PACK
(M3863A)
TYPE 12
VDC 4.2 Ah lithium manganese.
Disposable, recyclable, long-life, primary cells.
CAPACITY Minimum
300 shocks or 12 hours operating time.
INSTALL-BY-DATE Battery is
labeled with an install by date at least 5 years from date of manufacture.
STANDBY
LIFE Defines
how long the battery will power the AED I standby operation within the standby
temperature range (one battery insert test and no uses). 4 years minimum when battery is installed by
the Install-By-Date. 5 years typical.
ECG DISPLAY
(M3860A)
SCREEN High
Resolution LCD with bright backlight.
SCREEN
DIMENSIONS 2.8
inches wide x 2.3 inches high (7.0 cm x 5.8 cm)
DISPLAY
RANGE Differential:+/-
2 mV full scale (nominal)
SWEEP
SPEED 23
mm/second (nominal)
FREQUENCY
RESPONSE 1 Hz to 20 Hz
(-3dB) (nominal)
SENSITIVITY 1.16
cm/mV 9nominal)
HEART RATE 30 to
300 beats per minute updated each analysis period during monitoring.
MONITORED
LEAD Anterior-anterior
(lead II) placement with defibrillation pads (M3860A only)
DEFIBRILLATION
PADS, CABLE, CONNECTOR
CONFIGURATION DP2: Two pair of DP6: Six pair
HOW
SUPPLIED Disposable,
self-adhesive defibrillator pads with integrated cable and connector.
SURFACE
AREA 100
cm2 each (nominal active surface area)
CABLE
LENGTH Approximately
122 cm (48 inches)
MEDICAL
CONTROL/RECORDING FEATURES
STANDARD
EVENT REVIEW Elapsed time
and the number of shocks delivered are displayed on screen.
ENHANCED
EVENT REVIEW Optional Data
Card expands the above on-screen event review capabilities: Review chronological events in detail
including ECG
OPTIONAL
PC DATA CARD (M3854A)
Review
the patient's initial ECG on the screen
4
Hours of event & ECG recording or 30 minutes if Voice Recording is enabled.
ENVIRONMENTAL/PHYSICAL
REQUIREMENTS
SEALING Meets
IEC529 class IP54; with battery and data card tray installed
TEMPERATURE Operating: 0-50 Celsius (32-122 F)
HUMIDITY 0%
to 95% relative humidity (non-condensing)
STANDBY
TEMPERATURE 0-43
Celsius (32-109 F)
STANDBY
HUMIDITY 0% to
75% relative humidity (non-condensing)
Standby
applies to AED with battery installed and stored with defibrillation pads.
ALTITUDE -500
to 15,000 feet per MIL-STD-810E 500.3
Procedure
II
SHOCK/DROP
ABUSE MIL-STD-810E
516.4 Procedure IV
TOLERANCE (1
meter, any edge, corner or surface)
VIBRATION MIL-STD-810E
514.4-17
EMI
Requirements Tested
to EN60601-1-2 Limits.
EMI
(Radiated) Method
EN55011: 1998 Group 1 Level B.
EMI
(Immunity) Method
EN61000-4-3: 1998 Level 2.
Aircraft: Method RTCA/DO-160D: 1997 Section 21 (Category M-Charging).
AUTOMATED
AND USER ACTIVATED SELF TESTS
AUTOMATIC
SELF TESTS Tests
internal circuitry, waveform delivery system, battery capacity. Verifies
calibration of key circuits monthly.
AUTOMATIC
SELF TEST Daily when
stored within operating environmental
FREQUENCY conditions.
STATUS
INDICATION Dynamic
visual and audible indication of self test results. Indicates device readiness.
BATTERY
INSERTION TEST Upon
battery insertion, extensive automatic self tests and user interactive tests check
device readiness. Verifies calibration
of key circuits.
AUTOMATIC
STANDBY Instrument automatically monitors temperature
and
TEMPERATURE
MONITORING warns user if device is
stored outside of standby temperature range.
TRAINING
AND ADMINISTRATION PACK (M3864A)
FUNCTION Places
FR2 in scenario-based training mode. 10
real world scripts provided. Disables
energy delivery.
Permits
modification of preprogrammed FR2 protocol.
TYPE 12
VDC 1.1 Ah Rechargeable Nickel Metal Hydride
CAPACITY Min
4 hours training time
RECHARGE
TIME 90
minutes to full capacity using Charger, M3855A (sold separately)
CHARGER
(M3855A)
For
use with the Training and Administration Pack, M3864A.
POWER
INPUT/MAINS 100-240
VAC 47-63 HZ
POWER 30
Watts
CODERUNNER
WEB SOFTWARE
OPTIONAL Software
for review of data from optional PC Data Cards.
Refer to
FR2 User's Guide for detailed product use instructions.
All
specifications are based on 25 Celsius unless otherwise noted. Specifications
are
subject to change.
The FR2
and accessories are made of latex-free materials.

![]()